Specifications

13 
9.  Instrument: On the LC-20AD liquid chromatograph, open the drain by turning the knob 
anti-clockwise, and press purge. When purging is complete, close the drain by turning the 
knob clockwise 
10. Instrument: Press pump to turn on the pump and allow for pressure to stabilize and record 
the final pressure (allow solvent flow for about ~ 10 min and monitor pressure) 
Setting parameters for the analysis 
11. Software: Click on Data Acquisition tab and set parameters such as LC stop time, Pump 
mode, Solvent compositions, Detector wavelengths, Oven temperature etc. Save method 
by clicking on File>Save method file as> 
Sample preparation and using the autosampler 
12. Instrument: Filter sample through a Nylon 0.45 micron membrane, place sample in blue-
capped Prominence vial (minimum sample level has to be above the middle white marker 
on the vial), insert samples in the sample tray by opening the autosampler door (left 
bottom) and pulling out the tray—note the vial position number of each sample on the 
tray. Reposition the tray by pushing it in—make sure it clicks in, and close the 
autosampler door 
Starting the analysis 
13. Software: Click on Batch Table tab and enter sample information—Vial number (from 
the autosampler tray), Sample name, Injection volume…etc. To choose the previously 