Specifications

17 
Figure 3 – system is closed        Figure 4 – system is open to vacuum 
10. After 30 minutes, switch the system to argon for 5 – 10 seconds (should look like figure 
1). 
11. Repeat steps 8 – 10. 
12. Repeat step 8 and 9. 
13. Close the system by turning the bottom knob to its horizontal position (should look like 
figure 3). 
14. Switch the top knob to argon (should look like 
figure 2). 
15. On the flask, unscrew the black screw knob so 
the bulb and the top joint are now open to each 
other. 
16. SLOWLY open the middle solvent knob (figure 
5). (*NOTE* - as indicated by figure 5, the 
solvent knob does not need to be fully opened! 
You only need to have a steady flow of solvent). 
17. Fill flask about halfway and close the middle 
solvent knob by returning it to a horizontal 
position (should look like figure 2). 
18. Finally, open the flask to argon by turning the 
bottom knob to a vertical position (should look 
like figure 1). 
 Figure 5 – system is open to solvent 
Argon 
Vacuum 
Argon 
Vacuum 
Argon 
Vacuum 