Specifications

20 
  Concentration range (Low to High) 
  Recording range (Low to High) 
  Sensitivity 
  Auto scan (if necessary) 
  Response time (0.02) 
  Repetitions 
  If using Mulitpoint Curve, select order of curve and zero intercept 
Select OK to validate 
Prepare standards (at least 4). 
Auto zero the instrument with sample compartment empty or with a cuvette with solvent used as 
a blank. 
Select STANDARD push button. 
Place first standard in cuvette and put into the sample compartment. Select READ. The EDIT 
STANDARD box appears. Enter the appropriate data for the concentration and select OK. 
Repeat for each standard. 
Place unknown in cuvette and place in sample compartment. Select the UNKNOWN button and 
then the READ button for each unknown. 
When completed reading unknowns, select FILE, Save Channel. Click on the STANDARD box 
and type in a file name. Also type in comments if necessary. This will be saved as a .std file (do 
not type in the extension). 
Click on the UNKNOWN box and enter a file name (do not enter an extension). This will be 
saved as a .qnt file. 
Select OK to save files. 
TIME COURSE 
Acquire menu—Time Course 
Set parameters 
  EX / EM wavelength 
  Recording range (low to high) 