Specifications

3 
Asylum AFM SOP
1
Model: 
Serial Number: 
MFP Controls 
Before flipping over Housing, raise the stage to its MAX as to not crash the tip into the surface 
1.  Flip over Housing; make sure Housing is level 
2.  Click Live Video (Screen Image on bottom) 
3.  Turn on light at computer. Align at instrument to find the cantilever 
4.  Adjust back & right dials to increase SUM to its MAX VALUE 
5.  Adjust left dial to make REFLECTION = 0 
6.  Turn light off, close live video 
Tune Tab 
1.  Auto Tune High = 400 KHz (Vistaprobe) / 100 KHz (Asylum) 
2.  Target % = -5% 
3.  Click Auto Tune 
Main 
1.  Scan Size = 50 μm (can vary) 
2.  Scan Points = 256 
3.  Scan Lines = 256 
4.  Set Point = 900 (or 800) 
5.  Data Scale = 400 nm 
Meter 
1.  Click Engage 
2.  AMPLITUDE in range of 1 
3.  Z VOLTAGE 150 
4.  Bring down tip 
5.  AMPLITUDE gets to 0.90 (Ding!) 
6.  Go down until Z VOLTAGE = 30-40 
7.  Close Housing 
Main 
1.  Click Set Point Dot (Activates Wheel) 
1
 9/15/2010 