Under Template, select the template you would like to run for your first TGA sample. 
Templates are located in My Documents > My QUADERA > QMG220 > Templates. 
  A premade template is available for general use: Kevin Unknown 
Bargraph Cycle. This template will measure from 0-300 mass units over 
a span of 30 seconds (10 ms/amu). If you do not know what will come off 
your sample, use this sample. 
  If you have run an Unknown Bargraph Cycle or have an idea of the 
fragments coming off of your sample, an MID (Multiple Ion Detection) 
template can be used. This will measure up to 7 ions over the course of the 
TGA run and monitor their intensity differences over time. MID’s can be 
edited and saved in the QUADERA software in Window 2. Once in 
Window 2, select PrismaPlus and choose Kevin MID. On the bottom, 
click on the Edit tab. Only change tha Task Name and Mass columns 
then save as a new template. 
Type a name for your sample. Make sure both Filament and SEM are checked. Click 
Update Task.  