5.  Data Collection 
Click [Baseline] on the Photometer Button bar at the bottom of the screen and click [Ok] in the 
Baseline Parameters dialog box. Remember, Baseline corrects over the entire selected 
wavelength range. The Autozero function corrects at one wavelength. 
Perform a spectral scan by inserting the sample into the spectrophotometer. Click the [Start] 
button. After the scan, enter a file name in the New Data Set dialog box and click [Ok]. 
6.  Data Saving 
The data saving structure for UV Probe is flexible. Indepth instructions are available in 
Appendix B. For a simple way to save the acquired data, Select File > [Save As] and select the 
appropriate data directory in the [Save In] box at the top of the dialog box. Enter the correct file 
name, select .spc in the [Save As Type] list, and click [Save]. 
7.  Data Display 
If data has not been collected, open sample data. The suggested file is Rosebengal.spc. If you 
have only collected one spectra, you may want to open other demo data to show all the 
functionality of the software. The Graph pane on the right of the screen has 3 tabs. The Active 
tab shows one single spectrum. This is the data that will be manipulated. 