User's Manual

33
Warning(Avertissement)
To avoid risk of electric shock, this equipment must only be connected to a supply mains with
protective earth.
Afin d’éviter les risques de choc électrique, ces équipements ne doivent être raccordés qu’à une
conduite d’alimentation avec une terre protectrice.
Be aware of significant risks of reciprocal interference posed by me equipment during specific
investigations or tretments.
Soyez conscient des risques importants d’interférence réciproque que présente mon équipement
au cours d’enquêtes ou de travaux particuliers.
Notice
Turn on the workstation before turning on the Detector.
Check each component’s identification label before installation.
Check if the cable connection is perfectly clicked in.
Peripheral equipments that are connected to the Detector such as Workstation or Generator also
must be suitable for EN 60601-1 or EN 60950-1.
Power supply is specified as a part of me equipment or combination is specified as a me system.
When installing the system, do not plug in the AC Adapter plug into aconsent where it is difficult
to unplug the AC Adapter plug.
In order to discconnect the AC power, unplug the AC Adapter plug from the consent.
This device should be only used with the system equipment and the workstation PC.
Detector may have malfunction caused by following reasons:
when the power of the Control Box(BT-CB02) is off and when the power cables (BT-EC01, BT-TC01,
and BT-SC01) which are parts of the detector and cotrol box are disconnected. Make sure that the
power is on for the Detector with all the cables properly connected in order to have a successful
installation of the detector into the system. If problem still remains, refer to the detector manual.
Raw Image Viewer S/W which is provided with BSD3543W(BT-DA22W/BT-DB22W) can verify the
connection to Generator. The raw images saved on PC is in image format that does not contain
inforamtion of shooting environment and the patient. It is recommended for the user to select
and use a console S/W which can perform image processing, DICOM, changing format, managing
patient information, and connecting to a PACS system, and etc.
The Detector is tested for the integration to a Generator and a viewingsoftware as below.
-Generator:
kV Range: 40~125kV, 1kV step (Optional 40~150kV)
mA Range: 10 to 500mA
Timer Range: 0.001 to 10 sec, 38 steps
-Viewing Software:
S/W name: Raw Image Viewer (Refer to System Rquirement from p.22 and
Description of Raw Image Viewer from p.33.)
Data format: raw
The battery pack charger is used in a separate space.