47
Status Bar displays the image size, the coordinates of the current mouse position, and the level values.
Item Descriptions
Width, Height
- Shows the size of the image currently displayed on the screen.
- The values in parenthesis mean the starting coordinates (left, top)
and the ending coordinates (right, bottom) of the displayed image.
[X, Y] - Displays the coordinates and level of the current mouse location.
WW, WL
- Displays Window Width and Window Level.
- You can change by dragging the mouse while pressing the right
button.
7
Calibration
General Principle
The Purpose of Calibration
The center of the non-calibrated image is brighter than the edge due to hill effect of X-ray exposure. Generally,
the intensity of X-ray flux at center region of exposed area is higher than surroundings due to the X-ray expose
like cone shape. A calibration process is used to compensate for this effect. Generally, call this as ‘Flat Field
Correction’ (Bright calibration).
Notation
Parameter Description
Dark
Image acquired without X-ray exposure
(Without an object)
Bright Acquired image with X-ray exposure
Object Bright image with object, will be calibrated
Gain Gain of imaging system, offset subtracted image
Offset correction Offset subtract (dynamic offset)