User's Manual

48
Gain correction Compensate gain variance of pixel
Notice
X-ray detector should be used at stable state within driving temperature range.
Acquire the X-ray images after power on and 30 minutes warming up is recommended in order to obtain
high quality images.
Calibration Steps
Calibration steps are described as below.
Step Descriptions
Preperation Prepare for Calibration.
Step 1 Acquire Dark Frame.
Step 2 Acquire Bright Frame.
Step 3 Generate Defect map& Gain map.
[Preperation]
1. Click Calibration under Image Acquisition.
2. Adjust the Calibration Area by inputting Margin.