User's Manual

53
window will be appeared for reacquisition.
When acquisition is completed properly, Acquisition completed [1-1] - Save completed [1-1] message
appears in the Logwindow, and after 10 seconds, acquisition for the Bright frame automatically repeats.
Notice
If acquired Bright Image is not a suitable lelvel for Calibration, the following pop-up windows appear. Click OK
and retry. If the level is below the standard, increase the X-ray dose, and if the level is above the standard,
decrease the X-ray dose, and reexpose.
Notice
The Bright frame acquisition failure message and what thdy mean
- Not Ready for Image acquisition
The Detector is not ready for an image acquisition.
It automatically retries in about 10 seconds.
- Image read fail
Image acquisition has failed. A pop-up appears to check whether reacquisition will be proceeded or not.
If reacquisitions keep on failing, cancel and check the connection status with the Detector.
[Step 3] Defect map& Gain map
Notice
After acquiring all 3 Bright frames, it automatically starts to generate Defect map and Gain map.
After generating Reference.raw file, Completed pop-up appears.
1. Click OK on the pop-up.