User's Manual
56
Mode Descriptions
AED
AED(Auto Exposure Detection) automatically detects X-ray exposure signal without a
direct connection between Detector and Generator.
Trigger
1. It sends Ready Signal to Detector and Generator.
2. When Detector is ready, it sends EXPOSURE OUT signal to Generator.
[READY IN (GeneratorDetector), EXPOSURE OUT (DetectorGenerator)]
3. When Generator receives EXPOSRUE OUT signal, expose the X-ray.
2. Click Acquisition.
3. Press Hand switch to expose X-ray.
The acquired images will be displayed onRawImageViewer.
Notice
▩ If Not ready for image acquisition message pops up, wait 10 seconds and try again.
▩ Whenyou fail to acquire images, check the errors in the following section:
- Network settings: Jumbo Frame and Host IP settings
- Ethernet cable connection status
- Firewall is off.
- Reboot PC and detector.
▩ After clicking Acqusition, if you want to cancel the process, click Cancel.
Caution(Attention)
▩ Unexpected external shocks may cuase malfuction When shooting in AED mode.
▩ Des chocs externes inattendus peuvent provoquer une défaillance.
▩ Do not apply external shocks nor move the object until the image acquisition completes after shooting.
▩ N’appliquez pas de chocs externes et ne déplacez pas l’objet tant que l’acquisition de l’image n’est
pas terminée après le tournage.