User's Manual

56
Mode Descriptions
AED
AED(Auto Exposure Detection) automatically detects X-ray exposure signal without a
direct connection between Detector and Generator.
Trigger
1. It sends Ready Signal to Detector and Generator.
2. When Detector is ready, it sends EXPOSURE OUT signal to Generator.
[READY IN (GeneratorDetector), EXPOSURE OUT (DetectorGenerator)]
3. When Generator receives EXPOSRUE OUT signal, expose the X-ray.
2. Click Acquisition.
3. Press Hand switch to expose X-ray.
The acquired images will be displayed onRawImageViewer.
Notice
If Not ready for image acquisition message pops up, wait 10 seconds and try again.
Whenyou fail to acquire images, check the errors in the following section:
- Network settings: Jumbo Frame and Host IP settings
- Ethernet cable connection status
- Firewall is off.
- Reboot PC and detector.
After clicking Acqusition, if you want to cancel the process, click Cancel.
Caution(Attention)
Unexpected external shocks may cuase malfuction When shooting in AED mode.
Des chocs externes inattendus peuvent provoquer une défaillance.
Do not apply external shocks nor move the object until the image acquisition completes after shooting.
N’appliquez pas de chocs externes et ne déplacez pas l’objet tant que l’acquisition de l’image n’est
pas terminée après le tournage.