Electrical Specification
The unit meets the Collateral Standards of Electromagnetic compatibility - Requirements and tests EN 60601-
1:2007(IEC 60601-1-2) the limits and methods of measurement of electromagnetic disturbance characteristics of
industrial, scientific and radio frequency equipment EN 55011 Group 1, Class A, Electrical Equipment is subject in
regard to the electromagnetic compatibility(EMC) and its special precautionary measure
The unit must in reference to the mentioned EMC-hints in the accompanying documents be installed and
operated. Portable and mobile RF - communicating system (such as cell phones) can have influence to Electrical
Equipment.
Guidelines for the operator to use the e-Re Model device in electromagnetic environments.
electromagnetic emissions
The e-Re Model device is intended for use in the electromagnetic environment specified below.
The customer or the user of the e-Re Model device should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment
RF emissions
CISPR 11
Group 1
The e-Re Model device uses RF energy only for its internal function.
Therefore, its RF missions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The e-Re Model device is suitable for use in all establishments, including
domestic establishments and those directly connected to the public low-
voltage power supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
electromagnetic immunity
The e-Re Model device is intended for use in the electromagnetic environment specified below.
The customer or the user of the e-Re Model device should assure that it is used in such an environment.
IMMUNITY test
IEC 60601 test
level
Compliance level Electromagnetic environment
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 2 kV for power
supply lines
± 1 kV for
input/output lines
Mains power quality should be that of a typical
commercial or hospital environment.