User's Manual

62
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV differential
mode
± 2 kV common
mode
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 %
U
T
(>95 % dip in
U
T)
for 0,5 cycle
40 %
U
T
(60 % dip in
U
T)
for 5 cycles
70 %
U
T
(30 % dip in
U
T)
for 25 cycles
<5 %
U
T
(>95 % dip in
U
T)
for 5 s
<5 %
U
T
(>95 % dip in
U
T)
for 0,5 cycle
40 %
U
T
(60 % dip in
U
T)
for 5 cycles
70 %
U
T
(30 % dip in
U
T)
for 25 cycles
<5 %
U
T
(>95 % dip in
U
T)
for 5 s
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the e-Re Model device requires continued
operation during power mains interruptions, it is
recommended that the e-Re Model device be
powered from an uninterruptible power supply or
a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 0,3 A/m
If laser output distortion occurs, it may be necessary
to position the e-Re Model device further from
sources of power frequency magnetic fields or to
install magnetic shielding. The power frequency
magnetic field should be measured in the intended
installation location to assure that it is sufficiently
low.
NOTE 1
U
T is the a.c. mains voltage prior to application of the test level.
electromagnetic immunity
The e-Re Model device is intended for use in the electromagnetic environment specified below.
The customer or the user of the e-Re Model device should assure that it is used in such an environment.
IMMUNITY test
IEC 60601 test Compliance Electromagnetic environment