Service manual
Ultraview 1030/1050 Monitors — Service Manual
2-8
Inserting Recorder Paper
Two rolls of recorder thermal paper have been included with the recorder assembly option. To 
insert a roll of paper: 
1 Press the right side of the release bar at the top of the recorder assembly. The front of the 
recorder assembly is hinged at the bottom and will drop open, revealing two spoon-shaped 
arms that hold the paper roll spindle between them. 
2 Unroll a short length of paper from the roll, and orient the roll so that the paper feeds from the 
bottom. 
3 Slip the paper roll spindle between the plastic arms, and close the front of the assembly so that 
the end of the paper roll protrudes out of the recorder assembly just below the release bar. 
Power-ON Test
Each time the monitor is powered ON, internal system information is shown for approximately 
10 seconds and keys are displayed on the screen. The monitor is now ready for normal operation. 
External Devices
If an external SDLC device such as a Capnograph or other Flexport system interface is to be 
installed, the 9-pin connector on the rear of the remote module housing must be used. Only one 
external 9-pin port (the farthest from the monitor) can be used on a single bedside monitor. Even if 
multiple 9-pin SDLC ports on remote housings are available, only the one farthest from the monitor 
can be used for external devices. When this is the case, set the termination switch to OFF. 