Ultraview 1030/1050 Monitors — Service Manual
2-29
Setup
• Verify that the ECG waveform is displayed. 
3 Connect a waveform simulator to the pressure inputs.
4 Label and zero the channels.
• Verify that the numerics and waveforms are accurate. 
5 Touch RECORD, then touch one of the flashing parameter keys. 
• Verify that the recording is printed.
Note:
If the monitor is installed on a network, perform this test as described. Verify that the 
recordings are printed at the central printer.
Optional 90310 Wireless Network Interface (WNI)
90310 Front Panel
 
Rack mounting hardware — mounts the 90310 housing onto the housing pedestal mounts
Single housing kit (P/N 016-0365-00)
Double housing kit (P/N 016-0364-00)
 Power LED — when lit, indicates the presence of DC voltage to the 90310