Manual Chapter 1

1-19
Installing the Application
The following screen will be displayed.
8
10.··Select Yes and press the Enter key to continue.
Loading Firmware
Caution!
Reloading firmware in an UltraSpec 8000 analyzer resets the calibra-
tion factors. Always review calibration factors after loading firmware
to verify the settings are the same as in the factory supplied calibration
card. Refer to “Review Calibration” on page 2-32 for information
about calibrating the UltraSpec analyzer.
There may be up to a 20-30 second pause before loading begins. After
the firmware has been loaded, a list of application programs to be
loaded will be shown (if available on the same diskette). An error mes-
sage will appear if no application programs are located.
11.··If the error message occurred, insert the Application Diskette into the
computer drive A: (or B:); on the analyzer, press the Reset/Main key.
Otherwise, go directly to Step 12.
Loading Applications
A list of programs on the diskette will be shown.
12.··Select a program and press the Enter key.