7.2.9 Activate an Advanced Transient job
Activating a job lets you collect data for that job. It is similar to opening a file. Activate a job
when you want to change jobs.
Procedure
1.
From the Transient main menu, press F8 Job Manager.
2. Use the up and down arrow keys to select a job.
3. Press F3 Activate Job.
The Job ID appears at the top of the Transient main menu. You can now collect data
and save it to this job.
7.3 Sensors and inputs
7.3.1 Set the number of inputs
1. Activate a job.
2. From the Transient main menu, press F1 Setup / Acquire Data > F12 Input Setup > F1
Select Input.
3. Select one or more inputs.
Option Description
F2 Select/Unselect
Input A
Enable or disable data collection on input A.
F3 Select/Unselect
Input B
Enable or disable data collection on input B.
F4 Select/Unselect
Input C
Enable or disable data collection on input C.
F5 Select/Unselect
Input D
Enable or disable data collection on input D.
F7 A only Collect input data from input A. If you select this option, the inputs
on the left side of the screen automatically update.
F8 A and B Collect input data from A and B simultaneously. If you select this
option, the inputs on the left side of the screen automatically update.
F9 Triax Collect data from inputs A, B, and C. If you select this option, the
inputs on the left side of the screen automatically update. For the CSI
A0643TX triaxial accelerometer, the directions are relative to the
sensor:
Input A = Z direction
Input B = X direction
Input C = Y direction
4. Press Enter.
Advanced Transient
149