Option Description
F5 Set Sample
Time
Set the duration of the acquisition in seconds. The amount of memory
available for data storage may limit the number of samples. Changing the
sample time does not affect the sample rate. The default is 5. The Fmax also
affects the sample time. The maximum value is listed under the time.
4. Press Enter.
7.7.4 Enable PeakVue or Demodulation
1. Activate a job.
2.
From the Transient main menu, press F1 Setup / Acquire Data > F8 PeakVue Demod.
3. Select the desired option.
4. Press F10 Choose Filter to select a high-pass or band-pass filter.
Pre-defined frequency ranges are in the built-in demodulator. The analyzer adjusts
any entered frequency to the next highest predefined frequency.
5. Press Enter.
7.7.5 Enable or disable a tachometer trigger
1. Activate a job.
2. From the Transient main menu, press F1 Setup / Acquire Data > F9 Set Trigger.
3. Set the following options as necessary.
Option Description
F9 Set Trigger Type Enable or disable the tachometer trigger. The default is Disabled.
F11 Store Tach Times (Available only if Tach Trigger is On). Enable or disable tachometer time
values to save with the measurement point data.
4. Press Enter.
7.8 Collect transient data
Prerequisites
Attach the appropriate sensors to the analyzer and the equipment.
Procedure
1.
Create a job or activate saved a job.
2. From the Transient main menu, press F1 Setup / Acquire Data.
3. Ensure the data collection and display options are set as needed.
Advanced Transient
157