3. Press Enter.
3.5.2 Upgrade the firmware in the analyzer
Note
Updates to the analyzer programs are not included in the firmware upgrade. You must update the
programs separately from the firmware. Call technical support to see if any updates are available.
Prerequisites
Ensure the same connection type is set on the analyzer and in Data Transfer.
Procedure
1.
Copy the new firmware to a folder on your computer.
2. Connect the analyzer to the computer using the USB cable.
3. On the computer, open Data Transfer.
4. In the Options pane, set the Key Table and Firmware fields to the firmware folder
you created in step 1.
a. Click the Options icon.
b. Click Browse next to the Key Table field and select the folder containing the
firmware.
c. Click Browse next to the Firmware field and select the folder containing the
firmware.
d. Click Apply.
5. Turn off the analyzer.
6. Press and hold the Home key and then quickly press the Power key on the analyzer.
The CSI Special Functions Menu appears.
7. Press F2 Update Device Firmware via the USB Port.
The firmware upgrade begins. Follow the prompts. If the analyzer does not make the
connection, turn off the analyzer and go back to step 1.
3.5.3 Add or upgrade the programs in the analyzer
You can add or upgrade multiple programs at one time.
Prerequisites
Ensure the same connection type is set on the analyzer and in Data Transfer.
Update the firmware, if an update is available. You may see an error message if you
try to update or add programs before the firmware is updated.
Transfer files with CSI 2140