5 Installation of Intraskan dc
 5.6 Ground Connection Check:
Using a Multimeter, check the continuity between the following points:-
• Between Ground point of power board and Tube Head inner Cone metal part,
• Between Ground point of power board and Scissor Arm Guide rod and
• Between Ground point of power board and Base Unit Wall mounting plate as shown in Figure 28 .
• If any of the check fails then check the Ground connections inside Base Unit & Tube Head for Cable fault.
Figure 28: Check the continuity 
    of ground connection.
1
2
3
 5.7 Remote console Installation:
• Take the remote console template from packing box and fix it at the desired location on the stud. 
• Confirm the level of the template using spirit level. Mark the locations for the 2 mounting holes on the 
template and mark the wiring hole also at the locations shown in Figure 29 .
• Drill the 2 mounting holes marked using 5/32" drill bit to the depth of 1.26" and drill the direct wiring 
hole(>1.5" dia). 
Figure 29: 2 
Mounting holes
Remote Console Template