User's Manual

11
When the device prompts to calibrate the clock after the
self-test is completed, users need to set time themselves.
3.3 Start Test
Preparation before test:
Urine test strips that matched the device;
Urine sample that stored no more than 4 hours;
Absorbent paper for sucking residual urine;
Protective gloves for preventing contamination.
After the test strip is immersed in the sample, absorb the excess
sample liquid in both sides of test paper with absorbent paper, then put
the test strip flat on the device test paper tray. Keep the top of test strip
is aligned with the top of test paper tray.
The main interface displays test countdown, test sample number, user
name and the name of the test item.
According to the need, operate as follows:
: Short press this button once, the sample number add 1; long
press this button for 3 seconds, the paper tray out of the storehouse.
: Short press this button once, the sample number reduce 1;
long press this button for 3 seconds, the paper tray back to the
storehouse.
: Short press this button once to switch testing user; long press
this button for 2 seconds to enter the device setup interface; during the
test, short press it to exit.
: Short press this button once to begin 60-seconds countdown,
enter the test after the countdown return to zero; short press it once
again before zeroing, the countdown will directly return to zero and
immediately enter the rapid test status.
: Long press this button for 4 seconds, the device will power
off.
Test steps:
1) Place the test strip;
2) Under the main interface, press OK button
to begin testing the
current sample;