User's Manual

23
When installing, insert the clean tray into the device from underside,
hold the paper tray gently with hand, at this time, press the
direction
button for three seconds, then the paper tray will move
back to the device storehouse.
4.3 Clean the Paper Tray
Figure 4-1 Test paper tray
For daily cleaning, use the soft cloth dipped with distilled water or
absorbent paper to wipe the paper bracket and the calibration
comparator block, and make sure there is no dust, substance, nick, if
any found, please replace it with dealer.
If there is urine alkali in the paper bracket, use cotton swab dipping
with NaOH (concentration: 0.1mol/L) to wipe the paper bracket, and
use absorbent paper to wipe.
Note
Please do not clean with any substance that may scrape the paper
bracket and the calibration comparator block.
Calibration comparator block
Paper bracket
Drive rack
Positioning magnet