User's Manual

24
Please do not clean the calibration comparator block with any
solvent.
Please do not contact the calibration comparator block with
NaOH.
4.4 Disinfection
1) According to one of the following three methods to configure
disinfectant:
2% glutaraldehyde solution;
0.05% sodium hypochlorite solution -----1:100 dilution: add 1mL
sodium hypochlorite solution(concentration:5%) to 99ml water;
Isopropanol (70% -80%), without dilution.
2) Inject the disinfectant into a tall and narrow container for about
10 cm high.
3) Immerse the paper tray to the disinfectant, and keep the
calibration comparator block on the surface.
4) Soak for 10 minutes, then take it out and wipe it with absorbent
paper.
4.5 Waste Disposal
According to local regulations about biohazard waste disposal to
discard the waste generated during use.
4.6 Troubleshooting
When there is a fault with the urine analyzer or some functions can
not be achieved caused by users' improper operation, the urine
analyzer will displays error message, all error messages are as
follows:
Error message Solution
Abnormal
system
information
The memory has problems, the analyzer can
not properly read the system parameters, if it
still displays abnormal message after restarting
please contact the dealer.