Operation Manual
7
6.1.4
6.1.4
6.1.4
6.1.4 Pause
Pause
Pause
Pause alarm:
alarm:
alarm:
alarm:
A Alarm includ es the alarm of measure data's going beyond the limits, the alarm of low- power ,
and the alarm of finger out.
B I n the measuring interface,
if
the alarm function is on, during the period of alarming, alarm can
be suspended by short pressing the button, but the function will be renewed
in
about 60 seconds.
C If you want to turn off the alarm for good, you should enter the menu for operation.
6.1.5
6.1.5
6.1.5
6.1.5 Menu
Menu
Menu
Menu operations:
operations:
operations:
operations:
In the measuring interface ,the display direction can be changed by pressing the button with a short
push (click) . T here are four modes of data
di
splay that can be viewed.
Press the button with a prolonged push (1 second) to enter the Settings Menu I nterface (see Figure
5 ). Please
Please
Please
Please Note:
Note:
Note:
Note: When the display direction is lengthways , you can not enter the main menu
interface.C lick the power button to switch to landscape orientation.
The user can setup the following content
in
the Settings M enu
:
B acklight Brightness , A larm
setting,ID setting , data stor age ( R ecord) , turning on/off Bluetooth . T he specific operation methods
are as follow s
:
Please note
in
the Settings Menu:
CLICK
CLICK
CLICK
CLICK = short p ush of button and PRESS
PRESS
PRESS
PRESS = prolonged p ush of power button (1sec)
Figure 5 Main Menu Interface
A
A
A
A Backlight
Backlight
Backlight
Backlight adjustment
adjustment
adjustment
adjustment
I n the main menu interface, c lick the button to select “ Brightness ” , P ress the power button and hold
to adjust the backlight br ightness.
B
B
B
B Alarm
Alarm
Alarm
Alarm setting
setting
setting
setting
I n the main menu interface, c lick the power button to select “ Alarm ” , P ress the power button (1sec)
to enter the alarm setting interface as shown
in
F
i
g ure 6
:
a)
a)
a)
a) Adjust
Adjust
Adjust
Adjust ing
ing
ing
ing the
the
the
the high
high
high
high and
and
and
and low
low
low
low limit
limit
limit
limit s
s
s
s of
of
of
of alarm
alarm
alarm
alarm s
s
s
s
Click button to change the position of menu selection bar,and press button to set menu item.Move
the selection bar to “ Dir ” ,and press button to choose value adjusting direction
:
up or down.By
pressing button operation,the user could adjust the value of SpO
2
high limit (SPO
2
ALM HI), SpO
2
low limit (SPO
2
ALM LO), Pulse rate high limit (PR ALM HI), Pulse rate low limit (PR ALM
LO) .The alarm low limit can't be beyond the alarm high limit.The SpO
2
alarm range is
0%~100% ,the pulse rate alarm range is 0bpm~254bpm .