User's Manual
Do  not  use  the  cuff  in  the  area  where  the  treatment  is  being  performed  inside  blood  vessel  or  the  arteriovenous 
connection. This may cause temporary blockage of blood flow and cause injury to the patient. 
Do not use the cuff on the side of the mastectomy. 
When using the cuff to pressurize, some of the body’s functions may temporarily weaken. Do not use the measurement 
medical electrical equipment at the appropriate arm position. 
Do not move during measurement, it will have a delayed effect on the patient’s blood flow. 
The device need to be placed for 2 hours from the minimum storage temperature to being ready for its intended use. 
The device need to be placed for 4 hours from the highest storage temperature to being ready for its intended use. 
  Note   
The following conditions may also cause changes in the blood pressure measurement value. 
Take the measurement in one hour after meal or after drinking alcohol, coffee or after smoking, exercise, bathing; 
Using incorrect posture such as standing or lying down, etc; 
The patient speak or move his body during measurement; 
When measuring, the patient is nervous, excited, emotional instability; 
The room temperature rise or fall sharply, or the environment of measurement often changes; 
Measuring in a moving vehicle; 
The high and low location of cuff will cause changes in measurement results; 
Continuous measurement for a long time. 
7.2 Applying the Cuff 
Both left and right arm can be measured. 
Bare your arm or cloth close-fitting clothing during measurement. 
Carry out the operation in a room with comfortable temperature. 
When measuring, take the thick clothes off instead of rolling up the sleeves. 
In order to measure accurately, pay attention to applying the cuff properly (left arm). 
① Insert the arm cuff air plug in the cuff socket of sphygmomanometer. 
② Stretch cuff into a barrel for the arm can conformable enter into the barrel 
③ Left arm penetrate through the cuff, the air tube of the cuff will pass the top of your palm. 
④ Wrap the cuff to your upper arm. Make the air tube inside the forearm and aligned with your middle finger. 
⑤ The bottom of the cuff should be approximately 2cm~3cm above your elbow. 
⑥ Be fixed with cloths, and wrapped tight cuff, the arm and the cuff should not have gaps. 
① ②  ③ 
④ ⑤  ⑥ 
7.3 Measurement BP 
①Under “OFF” state, press “START/STOP” button to start measuring. 
During measurement, please keep correct pose and quiet state, the body could not move.The “Movement” icon appears 
if patient moves, and continue measuring may lead to inaccurate measurement. 
If you want to abort the measurement   
Press【START/STOP】button, the device will stop inflating, and release the air from the cuff. 
Display  the  measurement  results  after  finishing  measuring.The  pressure  bar  on  the  right  side  visually  demonstrates  the 
pressure level. 
7.4 Confirm the Measurement Value 
①The World Health Organization has established globally accepted standards for the assessment of hypertension readings.(In the 
clinic environment) 
Blood pressure level 
Systolic pressure 
Diastolic pressure 
Normal 
Pressure<130mmHg 
Pressure<85mmHg 
Normal Systolic Value 
130mmHg≤Pressure≤139mmHg 
85mmHg≤Pressure≤89mmHg 
Mild Hypertension 
140mmHg≤Pressure≤159mmHg 
90mmHg≤Pressure≤99mmHg 
Moderate Hypertension 
160mmHg≤Pressure≤179mmHg 
100mmHg≤Pressure≤109mmHg 
Severe  Hypertension/High  Blood 
Pressure 
180mmHg≤Pressure 
110mmHg≤Pressure 
Pressure bar at the right side 
The SYS higher than 135mmHg or the DIA higher than 85mmHg are used as the criteria of hypertension(In the home environment), 
and the pressure bar at the right side lights up in red. 
The SYS lower than 135mmHg and the DIA lower than 85mmHg are used as the criteria of normal pressure, and the pressure bar 
at the right side lights up all in green. 
The number of lights represents the blood pressure range. 
*Self-diagnosis and treatment using measured results may be dangerous. Follow the instructions of your physician. 
7.5 Upload date(optional for devices with Bluetooth function) 
①The data stored can be uploaded to master device by Bluetooth 
Time sync icon lights up if the device time is not synced, and the measured data cannot be uploaded to the terminal 
equipment via the Bluetooth. After syncing the time by Bluetooth connection of the device, Time sync icon goes off, 
and the measured data can be uploaded 
③After uploading the data to the master device, the local data will be deleted. 
  Note   
Wait at least 4-5 minutes between measurements. 
  When repeatedly measuring, the accurate blood pressure value may not be measured due to congestion in the arm. 
Please measure after the blood flow is smooth. 
  When the screen displays Err, the measure can't be carried out correctly. 
  Irregular pulse icon is  displayed  in the  measurement results if the pulse internal is irregular during measuring, 
which may cause it  is  unable to take measurement correctly. Please keep quiet and remeasure. If the irregular 
pulse icon appears frequently, please consult a doctor. 
  The minimum value of the patient’s  physiological signal is the minimum limit that the device can measure. The 
device may obtain inaccurate measurement results when it is operated below the minimum amplitude or minimum 
value of the patient’s physiological signal. 
*The device  will automatically turn off after five minutes in which  there is no operation to the  device,  even if you 
forget to turn the power off. 
Chapter8 Memory Function 
The device can store NIBP values automatically, display up to 199 set of measurement results. 
If 199 set of measurement data have been stored in current device, when saving the 200th set of data, the earliest set of 
data will be overwritten. If no measurement values, the memory values can be not numerated. 
Memory function can not be used during measuring. 
When there is no measurement values, “---” will display on the review interface. 
8.1 Review the Memory Value 
1.Under “OFF” state, press “M” button to display the average value of the latest three set of data, when the number of 
measurement data is less than three groups, it will supplement automatically. Continue to press “M” button in current 
interface to view all measurement records. 
8.2 Delete Memory Values 
1.Users can delete all memory values of the current user instead of separately delete one memory value 
2.Under  the  memory  interface,  press  “M”  button  and  “START/STOP”  button  simultaneously  for  more  than  5  s,  after 
“DEL” appears on the screen, all memory values will be deleted. 
  Caution   
When querying the measurement records, please press “M” button continuously to query one by one. 
Chapter9 Key and Symbols 
Your device may not contain all the following symbols. 
Signal 
Description 
Signal 
Description 
Attention! Please refer to the accompanying 
document (the user manual). 
Attention! Please refer to the accompanying 
document (the user manual). 
SYS 
Systolic pressure 
DIA 
Diastolic pressure 
MAP 
Mean blood pressure 
PUL 
Pulse rate (bpm) 
IP20 
Enclosure protection grade 
EMC 
Electromagnetic compatibility 
  Recyclable 
P/N 
Material code of manufacturer 
Batch code 
Use by date 
This way up 
Fragile, handle with care 
Keep dry 
Storage atmospheric pressure limitation 
Storage temperature limitation 
Storage humidity limitation 
Manufacturer 
Date of manufacture 
Batteries Power   
SN 
Serial number     
Flating 
Deflating 
Waste disposal mark, this symbol indicates 
that the waste of electrical and electronic 
equipment can not be disposed as an 
unclassified municipal waste and must be 
recovered separately. 
This item is compliant with Medical Device 
Directive 93/42/EEC of June 14, 1993,a directive 
of the European Economic Community. 
Class II equipment 
Type BF applied parts 
European Representative 
Irregular pulse 
Socket for power adapter 
Interface for connecting cuff 
Voice closed 
Voice enabled 
Large movement during measurement 
Cuff tied properly 
Artery indicator label 
Bluetooth 
time sync icon 
Chapter10 Error Message   
When the high pressure position appears "Err" and the low pressure position appears the error number, the measurement 
is not normal. 
Error Mark 
Causes   
Solutions 
Err2 
Err15 
Function abnormal 
Please contact us 
Err4 
Low battery 
Please replace the battery or link adapter 
Err6 
The cuff is not wrapped correctly. 
Wrap the cuff correctly (refer to Chapter 7) 
Err7 
Cuff leakage 
Replace with a new cuff 
Err8 
Air pressure error 
Keep arm, body still, measure again 
Err9 
The pulse signal is too weak or the cuff is loose. 
Wrap the cuff correctly (refer to Chapter 7) 
Err10 
Out of measure range 
Keep arm, body still, measure again 
Err12 
Cuff is blocked or squeezed 
Wrap the cuff correctly (refer to Chapter 7) 
Err11 
Err13 
The signal amplitude is too big owing to the arm 
or body moving or other reasons when measuring 
Keep arm, body still, measure again 
Err16 
Err19 
It takes too much time 
Chapter11 Troubleshooting 
Abnormal 
Phenomenons 
Causes 
Solutions 
BP measurement 
values too high or too 
low. 
Cuff is not connected correctly. 
Correctly connect cuff.   
Talk or move arm in measurement 
Keep quiet and restart a measurement. 
The turnup close oppress the arm 
Take off the clothes, and restart a measurement 
No pressure 
Cuff leakage 
Buy a new cuff. 
The cuff windpipe is not correctly 
connected with cuff 
Correctly connect. 
Cuff not inflate 
Contact us. 
Cuff deflate in short 
time 
Loose cuff 
Correctly tangle cuff. 
It can not carry on measurement ,even if press the measurement 
button 
Return on the power and restart a 
measurement. 
Abruptly turn the 
power off in adding 
pressure 
No use for a long time, the batteries can be 
exhausted owing to the changed 
temperature 
Replace all four batteries with new ones. 
Hold the on/off button 
but can not start the 
device 
Batteries are worn 
Replace all four batteries with new ones. 
The battery polarities is reversed   
Check the battery installation for proper 
placement of the battery polarities. 
Cuff inflation start before press the measurement button 
Stop using the device and contact us. 
Cuff never deflation 
Stop using the device and contact us. 
Air pressure error 
Deflation error 
Pull out the cuff to deflate. Stop using the 
device and contact us. 
Others   
Keep arm, body still, measure again. 
No press value displayed or the value unaltered when cuff inflating 
Pull out the cuff to deflate. Stop using the 
device and contact us. 
Other phenomenon 
Switch on the power once again and restart an 
operation. 
Replace the batteries. 
If no, please contact us. 
Chapter12 Maintenance, Cleaning and Keeping 
*Please do obey the precautions and correct operating methods in this user manual. Otherwise, we will not responsible for 
any fault. 
  Warning   
Remove the batteries before cleaning. The accessories and main unit must be separated for cleaning. 
Maintenance is not allowed during device using. 
Do not squeeze the rubber tube on the cuff. 
  Caution   
  High pressure disinfection to the device and accessories is not allowed. 
  Do not let water or cleaning agent flow into the socket to avoid device damage. 
  Do not soak the device and accessories in liquid. 
  If any damage or deterioration of the device and accessories is found, please do not use it. 
Maintenance: 
  Clean the device and accessories regularly. It is recommended to clean them every one month. When the device or 
accessory gets dirty, use a dry and soft cloth to wipe. If they are very dirty, it is available to dip the soft cloth into 
water or mild detergent, and wring out, then use the cloth for cleaning. 
  The  device  shall  be  inspected  and  calibrated  regularly  (or  according  to  inspection  standard  of  hospital).  The 
inspection  can be carried out in appointed institutions, or  by  professional personnel or  contact us  for inspection. 
Under the setting interface, Press the “START/STOP” button  once, after “CAL” appears  on the screen, press  “M” 
button for more than 15 s to enter the static pressure interface. 
  Advice     
  Do not use gasoline, volatile oil, diluent, etc. to wipe the device. 
  Do not clean or wet the cuff. 
Storage: 
  Advice     
  Do not expose the device in direct sunlight for long time, otherwise the display screen maybe damaged. 
  The basic performance and safety of the device are not affected by the dust or cotton wool in home   
environment,while the device shall not be placed where with high temperature, humidity or dusty. 
  Aged cuff may result in inaccurate measurement, please replace the cuff periodically according to the user manual. 
  To avoid device damage, keep the device out the reach of children and pets. 
  Avoid the device close to extreme high temperature such as fireplace, otherwise the device performance may be affected. 
  Do not store the device with chemical medicine or corrosive gas. 
  Do not place the device where there is water. 
  Do not place the device where with slope, vibration or impact 
  Take the batteries out if the device is not to be used for three months or longer. 
Chapter13 NIBP Specification 
Name 
Electronic Sphygmomanometer 
The degree of protection 
against ingress of water 
IP20 
Display 
LED 
NIBP Specifications 
Measurement Method 
Oscillometric method 
Working mode 
Automatic 
Operation mode 
Continuous operation 
Pressure Range 
0~297 mmHg(0~39.6 kPa) 
Measurement range 
Pressure 
SYS:30~270 mmHg(4~36 kPa) 
DIA:10~220 mmHg(1.3~29.3 kPa) 
Pulse: 40~240/min 
Inflation 
160±5 mmHg(21.33±0.67 kPa) 
Overpressure protect 
297±3 mmHg(39.6±0.4 kPa) 
Resolution 
Pressure: 1 mmHg(0.133 kPa) 
Accuracy 
Static pressure: ±3 mmHg(±0.4 kPa) 
Error 
The BP value measured by the device is equivalent with the measurement value of 
Stethoscopy, perform clinical verification in accordance with the requirements in ISO 
81060-2: 2013, whose error meets the followings: 
Maximum mean error: ±5 mmHg 
Maximum Standard deviation: 8 mmHg 
Operating Temperature/ 
Humidity 
  +5ºC~40 ºC            .    15%RH~85%RH(no condensation) 
Transport 
Transport by general vehicle or according to the order contract, avoid pounded, shake and 
splash by rain and snow in transportation. 
Storage 
Temperature: -20 ºC~+55 ºC; Relative humidity: ≤95 %((no condensation)); No corrosive 
gas and drafty. 
Atmospheric pressure 
700 hPa~1060 hPa 
Power supply 
4 "AA" alkaline batteries, AC Adapter(AC, 100 V-240 V, optional) 
Rated current 
≤ 600 mA 
Battery life 
When  the  temperature  is  23  ºC,  limb  circumference  is  270  mm,  the  measured  blood 
pressure is normal, 4 "AA" alkaline batteries cab be used about 300 times. 
Main Unit Dimensions 
122*110*83mm   
Main Unit Weight 
300 gram(without batteries) 
Safety classification 
Class Ⅱ equipment (power supplied by power adapter)/Internally powered equipment 
(power supplied by batteries) 
Type BF applied part 
Service life 
The service life of the device is five years or 10000 times of BP measurement. 
Date of manufacturer 
See the label 
Accessories 
Standard Configure: 
Adult Cuff: limb circumference 22-32 cm (upper arm center)   
User Manual, four "AA" alkaline batteries 
Optional Configure: 
AC Adapter: 
Input: voltage: AC 100 V~240 V frequency: 50 Hz/60 H      Rated current: AC 150 mA 
Output: DC 5.0 V±0.2 V 1.0 A 
Extra large adult Cuff: the range of limb circumference is 18-26 cm (middle part of upper 
arm) 
the range of limb circumference is 22-30 cm (middle part of upper arm) 
the range of limb circumference is 22-43 cm (middle part of upper arm) 
the range of limb circumference is 32-43 cm (middle part of upper arm) 
Appendix 
Table 1:   
Guidance and manufacturer’s declaration –electromagnetic emission 
The device is intended for use in the electromagnetic environment specified below. The purchaser or the user of the 
device should assure that it is used in such environment. 
Emission test 
Compliance 
RF emissions CISPR 11 
Group 1 
RF emissions CISPR 11 
Class B 
Harmonic emissions 
IEC 61000-3-2 
Class A 
Voltage fluctuations/flicker emissions 
IEC 61000-3-3 
Applicable 
Table 2: 
Guidance and manufacturer’s declaration-electromagnetic immunity 
The device is intended for use in the electromagnetic environment specified below. The purchaser or the user of the 
device should assure that it is used in such environment. 
Immunity test 
IEC60601 test level 
Compliance level 
Electrostatic  discharge 
(ESD)   
IEC 61000-4-2 
±8kV contact 
± 15 kV air 
±8kV contact 
±15kV air 
Electrical fast 
transient/burst 
IEC 61000-4-4 
±2kV for power supply lines 
± 1 kV for input/output line 
±2kV for power supply lines 
Not Applicable 
Surge   
IEC 61000-4-5 
±1 kV lines to lines 
±2 kV lines to earth 
±1 kV lines to lines 
Not Applicable 
Voltage dips, short 
interruptions and voltage 
vatiations on power supply 
input lines 
IEC 61000-4-11 
<5%UT(>95%dip in UT) for 0.5 cycle   
40% UT(60%dip in UT) for 5 cycle   
70%UT(30%dip in UT) for 25 cycle   
<5%UT(>95%dip in UT) for 5 sec 
<5%UT(>95%dip in UT) for 0.5 cycle   
40% UT(60%dip in UT) for 5 cycle   
70%UT(30%dip in UT) for 25 cycle   
<5%UT(>95%dip in UT) for 5 sec 
Power  frequency  (50  / 
60Hz) magnetic field   
IEC 61000-4-8 
30 A/m 
30A/m 
Table 3:   
Guidance and manufacturer’s declaration – electromagnetic immunity 
The device is intended for use in the electromagnetic environment specified below. The customer the user of the 
device should assure that it is used in such environment. 
Immunity test 
IEC 60601 test level 
Compliance level 
Conducted RF 
IEC61000-4-6 
3 V 
0,15 MHz – 80 MHz 
6 V in ISM bands between 
0,15 MHz and 80 MHz 
3 V 
0,15 MHz – 80 MHz 
6 V in ISM bands between 
0,15 MHz and 80 MHz 
Radiated RF 
IEC61000-4-3 
10 V/m 80 MHz- 2.7 GHz 
10 V/m80 MHz- 2.7 GHz 
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 
reflection from structures, objects and people. 
a    Field  strengths from fixed transmitters, such as  base stations for radio  (cellular/cordless) telephones and land 
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with 
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey 
should be considered.  If the measured field  strength in  the  location  in which the the  device is  used exceeds the 
applicable RF compliance level above, the the device should be observed to verify normal operation. If abnormal 
performance is observed, additional measures may be necessary, such as reorienting or relocating the the device. 
Table 4: 
Guidance and manufacturer’s declaration - electromagnetic Immunity 
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the 
device should assure that it is used in such an environment 
Radiated RF 
IEC61000-4-3 
(Test 
specifications 
for 
ENCLOSUR
EPORT 
IMMUNITY 
to RF wireless 
communicatio
ns equipment) 
Test 
Frequency 
(MHz) 
Band a) 
(MHz) 
Service a) 
Modulat
ion b) 
Modulat
ion b) 
(W) 
Distan
ce 
(m) 
IMMUNIT
Y 
TEST 
LEVEL 
(V/m) 
385 
380– 
390 
TETRA 400 
Pulse 
modulat
ion b) 
18 Hz 
1,8 
0,3 
27 
450 
380– 
390 
GMRS 460, 
FRS 460 
FM c) 
± 5 kHz 
deviatio
n 
1  kHz 
sine 
2 
0,3 
28 
710 
704– 
787 
LTE Band 13, 
17 
Pulse 
modulat
ion b) 
217 Hz 
0,2 
0,3 
9 
745 
780 
810 
800– 
960 
GSM 800/900, 
TETRA 800, 
iDEN 820, 
CDMA 850, 
LTE Band 5 
Pulse 
modulat
ion b) 
18 Hz 
2 
0,3 
28 
870 
930 
1720 
1700–199
0 
GSM 1800; 
CDMA 1900; 
GSM 1900; 
DECT; 
LTE Band 1,3,4,25; 
UMTS 
Pulse 
modulat
ion b) 
217 Hz 
2 
0,3 
28 
1845 
1970 
2450 
2400–257
0 
Bluetooth,WLAN,802.11 
b/g/n,RFID  2450,LTE 
Band 7 
Pulse 
modulat
ion b) 
217 Hz 
2 
0,3 
28 
5240 
5100–580
0 
WLAN 802.11a/n 
Pulse 
modulat
ion b) 
217 Hz 
0,2 
0,3 
9 
5500 
5785 
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the 
ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3. 
a) For some services, only the uplink frequencies are included. 
b) The carrier shall be modulated using a 50 % duty cycle square wave signal. 
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not 
represent actual modulation, it would be worst case. 