User`s manual

22
Selftest and calibration check
If you feel that your programmer does not react according to your expectation, please run the
programmer (ISP connector) selftest using Diagnostic POD (Diagnostic POD for ISP
connectors #2), enclosed with the standard delivery package.
Selftest of programmer
Insert 48 pins diagnostic POD - type I into ZIF socket of the programmer. 48 pins
diagnostic POD - type Imust be inserted as 48 pins device.
Run selftest of programmer in PG4UW (Programmer / Selftest plus).
Selftest of ISP connector
Insert Diagnostic POD for ISP connectors #2 into ZIF socket of the programmer.
Diagnostic POD for ISP connectors #2 must be inserted as 48 pins device.
Interconnect 20 pins connector of Diagnostic POD for ISP connectors #2 with an ISP
connector of the programmer with an ISP cable, included in delivery programmer
package. Be sure that pins are interconnected properly (i.e. 1-1, 2-2, ..., 20-20).
Run selftest of ISP connector in PG4UW (Programmer / Selftest ISP connector…).
Calibration test
For optimal results with programmer we recommend you also undertake every 6 months an
extended test to check the calibration using 48 Pins Calibration test POD, Type I (optional
accessories, ord.no. 70-0438).
Insert 48 Pins Calibration test POD, Type I into ZIF socket of the programmer. 48 Pins
Calibration test POD, Type I must be inserted as 48 pins device.
Run calibration test of programmer in PG4UW (Programmer / Calibration test).