10.4 Specifications 
10.5 Channel Menu 
Each Strain-Gage Input channel is configured and calibrated via its channel menu. Detailed instructions on 
operating the instrument in the SETUP Menu mode can be found in “SETUP Menu mode”. A diagram of all menus 
is located in “Setup Menu Reference” in Chapter 18. 
10.5.1 OPERATION Sub-Menu 
This menu controls the operation of this channel when the instrument is in the RUN mode. 
AUTO-ZERO Menu Item 
This enables or disables the channel’s automatic zeroing function. The auto-zero function can be beneficial by 
removing the effect of slight temperature shifts in the transducer’s output. The two options for this menu item are 
“ON” and “OFF”. 
When enabled, the channel will reset its tracking value to zero when the tracking value has never been more 
positive than +10 or more negative than –10 display counts for any 10 second period. 
For example, assume that the DISPLAY. DECPT menu item is set to 000.00 (two decimal points). If the tracking 
value never leaves the range of –000.10 to +000.10 for any 10-second period, then tracking value will reset to 
000.00 automatically. 
To compensate for the transducer’s initial warm-up shift, the auto-zero function operates every 2 seconds during 
the first 60 seconds of RUN mode operation. 
TRANSDUCER INPUT
Transducer type  Full-bridge, strain gage 
Excitation Voltage  5 or 10 VDC, short circuit protected with sensing 
Transducer full-scale output  .5 to 21 mV/V @ 5V excitation 
.5 to 10.5 mV/V @ 10 V excitation 
Amplifier Gain Selection  Automatic 
Calibration Type  Shunt, mV/V or 2-,3- or 5-point known load 
Differential Input Voltage  +/- 105mV (max.) 
A/D Converter  24-bit Sigma-Delta 
Low-pass filter  Digital, 24-tap FIR 
Resolution and Frequency Response  See “FREQ.RESPONSE Menu Item” 
INSTRUMENT-ONLY ACCURACY 
(Sense wires used; Frequency Response setting <=16Hz; 
Linearity, repeatability & hysteresis) 
mV/V Calibration  ±0.1% F.S. 
Known-Load Calibration  ±0.01% F.S. 
AUXILLIARY INPUTS 
Quantity 2 
Type  Momentary contact closure 
Response Time  <5ms 
Field-Selectable Functions  Tare on, tare off, peak/valley clear, peak/valley hold, track hold. 
ANALOG OUTPUT 
Output voltage  5, ±5, 10 or ±10 VDC (field selectable) 
Output current range (optional current output 
channels) 
4-20mA 
Source  Any channel’s track, peak or valley value 
Isolation 500V 
Resolution 13 bits 
Frequency Response  Same as input when driven by the same channel’s tracking data 