FREQ.RESPONSE Menu Item 
This sets the frequency response, step response, and resolution of the channel according to the table below. The 
tracking value, peak/valley detector values and analog output on this channel are all affected. The default value of 
016. HERTZ is suitable for most applications. 
Frequency 
Response 
Step Response 
(ms) (typical) 
Resolution (counts) (not including 
min. 10% overrange/underrange 
capability) 
002. HERTZ/FAST  40  ±50000 
002. HERTZ  440  ±50000 
008. HERTZ  110  ±25000 
016. HERTZ  55  ±20000 
032. HERTZ  28  ±10000 
050. HERTZ  16  ±5000 
100. HERTZ  8  ±5000 
250. HERTZ  3  ±2000 
500. HERTZ  2  ±2000 
800. HERTZ  2  ±2000 
POWER-ON SOURCE Menu Item 
This menu selects which value is displayed by the channel when first entering the RUN mode. 
The choices are: 
•  “TRACK” means the live tracking value of the channel 
•  “PEAK” means the highest value of the channel. 
•  “VALLEY” means the lowest value of the channel. 
•  “TRACK/PEAK” means the live tracking value and the highest value of the channel (only available on DFI 
SC3004) 
•  “TRACK/VALLEY” means the live tracking value and the lowest value of the channel (only available on DFI 
SC3004) 
•  “PEAK/VALLEY” mans the highest and lowest values of the channel.  (only available on the DFI SC3004) 
10.5.2 DISPLAY SETUP Sub-Menu 
This menu controls how data values are displayed by the channel and transmitted via serial communications. 
DISPLAY.DIGITS Menu Item 
Selects the number of digits displayed by the channel. The choices are: 
•  “5 DIGIT-BIPOLAR” displays both positive and negative numbers with five full digits (±999999). 
•  “6 DIGIT-UNIPOLAR” displays positive numbers with six full digits (999999) and negative numbers with five full 
digits (-99999). 
•  “7 DIGIT UNIPOLAR” displays positive number with seven full digits (9999999) and negative numbers with six 
digits (-999999). 
This menu item may be automatically updated by a transducer’s Signature Module. 
DISPLAY.DECPT Menu Item 
Selects the decimal point location on the channel’s display and serial communications output. Use the [UP] and 
[DOWN] buttons to move the decimal point to the right and left. 
This menu item may be automatically updated by a transducer’s Signature Module. 
DISPLAY.COUNT-BY Menu Item 
Determines by what value the display increments or decrements. This will make the display less sensitive to a 
noisy signal. The choices are: 
• “00001” 
• “00002” 
• “00005” 