Manual
CF 66  28  Rev. C 2/05 
• “00010” 
• “00020” 
• “00050” 
• “00100” 
• “00200” 
This menu item may be automatically updated by a transducer’s Signature Module. 
DISPLAY.UNITS Menu Item 
Specifies the four-character label that is displayed to the right of the channel’s values. 
This menu item doesn’t change the mathematical scaling of the channel’s values; that can be changed by altering 
the “CALIBRATION DATA->FULL SCALE VALUE” menu item and then performing a re-calibration. 
When a character position is flashing press the [UP] or [DOWN] button to change the character. Press [ENTER] to 
advance to the next character. 
This menu item may be automatically updated by a transducer’s Signature Module. 
DISPLAY.AVERAGE Menu Item 
Controls the speed with which the channel’s display values will update. Display averaging does not affect the 
channel’s analog output or its peak/valley detection, which will proceed at the speed selected by the “OPERATION-
>FREQ.RESPONSE” menu item. The choices are: 
•  “ON” means that the display will update four times each second. The channel’s values will be averaged for ¼ 
second, then displayed. 
•  “OFF” means that the channel’s display will update as quickly as possible. 
10.5.3 AUXn FUNCTION Menu Items 
The AUX1 FUNCTION and AUX2 FUNCTION menu items determine what happens when the Auxiliary Function 
pins (labeled as “AUX1” and “AUX2”) on the channel’s connector are activated. These pins are “activated” when 
they are connected to the (-) Signature (labeled as “-MEM”) pin. The choices are: 
•  “DISABLED” means that activating the pin does nothing. 
•  “TRACK HOLD” means that the tracking, peak and valley values will not be updated. 
•  “HIGH/LOW HOLD” means that the peak and valley values will not be updated. 
•  “HIGH/LOW CLEAR” means that the peak and valley values are reset. 
•  “TARE ON” activates the channel’s Tare function (resets it to zero). 
•  “TARE OFF” deactivates the channel’s Tare function. 
As the Auxiliary Function pins are not isolated, it is recommended that a push-button switch or relay is used to 
connect these pins to the (-) Signature pin. 
10.5.4 CALIBRATION TYPE Menu Item 
This chooses the type of calibration technique to be used.  There are three methods that can be used to calibrate 
the transducer to the Input Channel. Each has advantages and disadvantages as described in the table below. 
Table 10-1: Comparison of Calibration Types 
Shunt Calibration  mV/V Calibration  2-, 3-, or 5-point Known 
Load Calibration 
RECOMMENDED 
…for most applications  No  Yes  No 
…when frequently swapping transducers  Yes  Yes  No 
…when best possible accuracy required  No  No  Yes 
PROS and CONS       
Relative accuracy  Good  Better  Best 
Requires actual, calibrated loads to be 
applies 
No No Yes 
Automatically calibrates with Signature  Yes Yes No 










