Manual
CF 66  44  Rev. C 2/05 
Step 3: Move the LVDT’s armature to the position that displays the lowest value of raw analog-to-digital converter 
readings (i.e. a number close to 00.000%). This is the LVDT’s electrical null. 
Step 4: Firmly clamp down the LVDT’s body. 
Step 5: Move the LVDT’s armature to roughly the position you wish to define as “positive full scale”. If the 
displayed analog-to-digital converter reading is a negative value, you must swap the wires connected to the 
(+)Signal and (-) Signal pins (labeled as “+SIG” and “-SIG”) on the channel’s 12-pin connector to make the reading 
positive. 
11.7 Analog Output Configuration 
11.7.1 Identifying the Output Type 
An AC/AC-LVDT Input channel is available with one of two types of digital-to-analog (DAC) outputs: voltage or 
current. You can determine which type of output a channel has by one of three ways: 
•  Consulting the instrument’s System Calibration Sheet. 
•  Examining the SYSTEM MENU->CONFIGURATION->CHANNEL nn TYPE menu item where nn is the number 
of the channel. If the channel’s type is AC-AC LVDT V, it has a voltage output. If the channel’s type is AC-AC 
LVDT I, it has a current output. 
•  Examining the channel’s circuit board as shown in the figure below. 
11.7.2 Channel Menu Items 
The Analog Output can be driven by any channel’s track, peak or valley value. 
See the “Channel Menu” section earlier in this chapter for a complete listing of SETUP menu items available on the 
DAC.SETUP sub-menu. 
11.7.3 Output Selection 
Jumpers located on the channel’s circuit board determine what outputs are generated when the value selected to 
drive the Analog Output (from the DAC.CHANNEL and DAC.SOURCE menu items equals the DAC. FULL SCALE 
and DAC.ZERO SCALE settings). 
Figure 11-2: Digital-to-Analog Output Jumper Locations 
  DAC.ZERO SCALE  DAC.FULL SCALE  J30  J31 
 Output Output jumper jumper 
CHANNELS WITH VOLTAGE OUTPUT 
0-5V  2.5 Volts  5 Volts  open  closed 
±5V  0 Volts  5 Volts  open  open 
0-10V  5 Volts  10 Volts  closed  closed 
±10V  0 Volts  10 Volts  closed  open 
CHANNELS WITH CURRENT OUTPUT 
4-20mA 4mA  20mA open open 
4-20mA 12mA  20mA open closed 