Manual
CF 66  51  Rev. C 2/05 
Figure 12-7: Excitation and Signal Jumper Locations on the High-Level Input Channel 
12.4 Calibration Procedure 
If you are not familiar with operating the instrument in the SETUP menu mode, see “SETUP Menu mode”. A listing 
of all menu items is given in “Setup Menu Reference”. 
Step 1: Wire the transducer to the channel’s connector. 
Step 2: Set the Excitation and Signal jumpers appropriate for the transducer, amplifier or DC-DC LVDT. 
Step 3: Enter the CALIBRATION TYPE. 
There are two methods that can be used to calibrate the transducer to the Input Channel. Each has advantages 
and disadvantages as described in the CALIBRATION TYPE menu item section.  It is important to know your 
application in order to select the appropriate calibration type. 
Step 4: Enter the CALIBRATION DATA 
Otherwise, consult the Certificate of Calibration for the transducer when entering information in the CALIBRATION 
DATA sub-menu. 
Step 5: Perform the calibration. 
Using the CALIBRATE menu item starts the calibration process. You will be prompted to apply loads to the 
transducer as required. 
12.5 Specifications 
TRANSDUCER INPUT   
Transducer types  Amplified with voltage or current output, (field selectable) 
Excitation voltage 
(1)
±15 VDC @ 40 mA max. 
+28 VDC @ 75 mA max. 
+12 VDC @ 40 mA max. 
(field selectable) 
Transducer full-scale output 
±1, ±5, ±10 VDC 
4-20 mA (field selectable) 
Calibration Type  Shunt, or 2-, 3- or 5-point known load 
A/D Converter  24-bit Sigma-Delta 
Low-pass filter  Digital, 24-tap FIR 
Resolution & Frequency Response 
See “FREQ. RESPONSE” Menu Item, Section 12.6.1 
INSTRUMENT-ONLY ACCURACY 
(Sense wires used; Frequency Response setting <=16Hz; Linearity, 
repeatability & hysteresis) 










