Manual
CF 66  67  Rev. C 2/05 
16.3 Setup Procedure 
If you are not familiar with operating the instrument in the SETUP menu mode, see “SETUP Menu Mode”. A listing 
of all menu items is given in the “Setup Menu Reference” in Chapter 18. 
Step 1: Examine the System Calibration Sheet. The instrument’s System Calibration Sheet contains important 
information about what each Mathematics Channel does and what Function Inputs and/or Limit Output pins of the 
System connector are used. 
Step 2: Wire to the pins on the System connector. 
See “System Connector” in Section 5.1. It is important to read the System Calibration Sheet to determine the 
wiring for your application. 
Step 3: Verify operation. 
16.4 Channel Menu 
Each Mathematics Virtual Channel can have its configuration and SensoCode programming examined via its 
channel menu. It is not possible to alter the SensoCode program with the SETUP menu. Detailed instructions on 
operating the instrument in the SETUP menu mode can be found in “SETUP Menu mode”. A diagram of all menus 
is located in “Setup Menu Reference” in Chapter 18. 
16.4.1 SENSOCODE P/N Menu Item 
This menu item displays the part number and revision level of the SensoCode program loaded in this Mathematics 
Virtual Channel. 
16.4.2 USER VALUES Menu Item 
This sub-menu contains menu items for each of the User Values. User Values are constant numeric values that 
can be edited by the user and read by a SensoCode program. The use of each User Value depends on the 
SensoCode program loaded in the Mathematics Virtual Channel. 
By default, the four items on this menu are USER VALUE 1, USER VALUE 2, USER VALUE 3 and USER VALUE 
4. However, a SensoCode program can alter the menu item text for each of the User Values to make them more 
relevant to the operator. For example, USER VALUE 1 might be re-titled JELLY BEAN COUNT. 
16.4.3 VIEW SENSOCODE Menu Item 
This menu item allows the display of each “step” (either an operand or an instruction) of the SensoCode program 
presently loaded into a Mathematics Virtual Channel. Press [ENTER] will display the next step of the SensoCode 
program until the entire program has been viewed. 
SensoCode programs cannot be altered with the SETUP Mode menu; they can only be loaded via the serial port 
with the Instrument Utility Software. 
16.4.4 VIEW COMMANDS Sub-Menu 
This sub-menu contains menu items for each of the Command Strings that can be used by a SensoCode program. 
Command Strings are text constants that can be read by a SensoCode program. They cannot be changed with the 
SETUP menu. The use of each Command String depends on the SensoCode program loaded in the Mathematics 
Virtual Channel. 
The items on this menu are COMMAND 0, COMMAND 1, COMMAND 2, COMMAND 3, COMMAND 4, COMMAND 
5, COMMAND 6 and COMMAND 7. 
16.4.5 DISPLAY SETUP Sub-Menu 
This menu controls how data values are displayed by the channel and transmitted via serial communications. 
DISPLAY.DIGITS Menu Item 
Selects the number of digits displayed by the channel. The choices are: 
•  “5 DIGIT-BIPOLAR” displays both positive and negative numbers with five full digits (±99999). 