Instruction Manual
Record your “INPUT1” and “READ 1” numbers, and your “INPUT2” and “READ 2” numbers, record them below and 
proceed as follows: 
 Example
INPUT1=  000003. 
READ 1 =    000000. 
INPUT2 =    100018. 
READ 2 =    068000. 
For the purpose of this example, we will use the numbers described above for the following procedure: 
PRESS  THEN 
PRESS (TO 
CHANGE IF 
REQUIRED) 
UNTIL 
DISPLAY 
SHOWS 
COMMENTS 
‘MENU’    “IN.SC.OF”  Press the ‘MENU’ button until the display shows “IN.SC.OF” Input 
Scale and Offset. 
‘MIN’    “INPUT1”  Press the ‘MIN’ button and the display will show “INPUT1”. 
‘MIN’    “XXXXXX”  Press the ‘MIN’ button and the display will show some 6-digit 
number. 
‘MAX’  ‘MIN’  “000003.”  Using the ‘MAX’ button to change the value of the flashing digit 
and the ‘MIN’ button to scroll to the next digit to the right, enter 
your “INPUT 1” number on the display. 
‘MENU’    “READ 1”  Press the ‘MENU’ button and the display will show “READ 1". 
‘MIN’    “XXXXXX”  Press the ‘MIN’ button and the display will show some 6-digit 
number. 
‘MAX’  ‘MIN’  “000000.”  Using the ‘MAX’ button to change the value of the flashing digit 
and the ‘MIN’ button to scroll to the next digit to the right, enter the 
engineering value that you want your “INPUT 1” number to display 
on the meter. 
‘MENU’    “INPUT2”  Press the ‘MENU’ button and the display will show “INPUT2”. 
‘MIN’    “XXXXXX”  Press the ‘MIN’ button and the display will show some 6-digit 
number. 
‘MAX’  ‘MIN’  “100018.”  Using the ‘MAX’ button to change the value of the flashing digit 
and the ‘MIN’ button to scroll to the next digit to the right, enter 
your “INPUT 2” number on the display. 
‘MENU’    “READ 2”  Press the ‘MENU’ button and the display will show ‘READ 2". 
‘MIN’    “XXXXXX”  Press the ‘MIN’ button and the display will show some 6-digit 
number. 
‘MAX’  ‘MIN’  “068000.”  Using the ‘MAX’ button to change the value of the flashing digit 
and the ‘MIN’ button to scroll to the next digit to the right, enter the 
engineering value you want your “INPUT 2” to display. 
‘MENU’    “STORED”  Press the ‘MENU’ button to store your selection and the display 
will momentarily show “STORED” then “DEC PT”. 
Now you are ready to position your decimal point position by completing the following steps: 
8.3.3 To Select Decimal Point Position (DEC PT) 
The following is the procedure for selecting your decimal point position. 
PRESS  THEN PRESS 
(TO CHANGE 
IF REQUIRED) 
UNTIL DISPLAY 
SHOWS 
COMMENTS 
‘MENU’    “DEC PT”  Press the ‘MENU’ button until the display shows “DEC PT”. 
‘MIN’  ‘MAX’  “FFF.FFF”  Press the ‘MIN’ button and the display will show “FFFFFF.”or 
the previously selected position. Press the ‘MAX’ button to 
select the decimal point position you require, the meter 
displays the previously selected 
CF 67  24  M1291/N/0403 11279ML-02 Rev. A 