Instruction Manual
PRESS  THEN PRESS 
(TO CHANGE 
IF 
REQUIRED) 
UNTIL 
DISPLAY 
SHOWS 
COMMENTS 
‘MENU’  “IN.SC.OF”  Press the ‘MENU’ button until the display shows “IN.SC.OF” Input 
Scale and Offset. 
‘MIN’  “INPUT1”  Press the ‘MIN’ button and the display will show “INPUT1”. 
‘MIN’  “XXXXXX”  Press the ‘MIN’ button and the display will show some 6-digit 
number. 
‘MENU’  “IN.SC.OF”  Press the ‘MENU’ button until the display shows “IN.SC.OF” Input 
Scale and Offset. 
‘MIN’  “INPUT1”  Press the ‘MIN’ button and the display will show “INPUT1”. 
‘MIN’ 
“XXXXXX”  Press the ‘MIN’ button and the display will show some 6-digit 
number. 
‘MAX’  ‘MIN’  “000000.”  Using the ‘MAX’ button to change the value of the flashing digit and 
the ‘MIN’ button to scroll to the next digit to the right, enter “000000.” 
on the display. 
‘MENU’  “READ 1”  Press the ‘MENU’ button and the display will show ‘READ 1". 
‘MIN’ 
“XXXXXX”  Press the ‘MIN’ button and the display will show some 6-digit 
number. 
‘MAX’  “000000.”  Using the ‘MAX’ button to change the value of the flashing digit and 
the ‘MIN’ button to scroll to the next digit to the right, enter “000000” 
on the meter. 
‘MENU’  “INPUT2”  Press the ‘MENU’ button and the display will show “INPUT2”. 
‘MIN’ 
‘MIN’ 
“XXXXXX”  Press the ‘MIN’ button and the display will show some 6-digit 
number. 
‘MAX’  “100000.”  Using the ‘MAX’ button to change the value of the flashing digit and 
the ‘MIN’ button to scroll to the next digit to the right, enter “000000.” 
on the display. 
‘MENU’  “READ 2”  Press the ‘MENU’ button and the display will show “READ 2". 
‘MIN’ 
‘MIN’ 
“XXXXXX”  Press the ‘MIN’ button and the display will show some 6-digit 
number. 
‘MAX’  ‘MIN’  “100000.”  Using the ‘MAX’ button to change the value of the flashing digit and 
the ‘MIN’ button to scroll to the next digit to the right, enter “100000.” 
on the display. 
‘MENU’    “STORED”  Press the ‘MENU’ button to store your selection and the display will 
momentarily show “STORED” then “DEC PT”. 
Proceed with the following steps for positioning your decimal point position: 
PRESS  THEN 
(TO CHANGE 
PRESS IF 
REQUIRED) 
UNTIL 
DISPLAY 
SHOWS 
COMMENTS 
‘MENU’    “DEC PT”  Press the ‘MENU’ button until the display shows “DEC PT”. 
‘MIN’  ‘MAX’  “FFF.FFF”  Press the ‘MIN’ button and the display will show “FFFFFF.” Or the 
previously selected position. Press the ‘MAX’ button to select the 
decimal point position you require, the meter displays the previously 
selected decimal point location within the “F’s”. 
‘MENU’    “CNT BY”  Press the ‘MENU’ button to store your decimal point selection and 
the meter will momentarily display “STORED” only if you have made 
a change and then “CNT BY”. 
You are now ready to scale your meter using Input Scale and Offset (“IN.SC.OF”) without connecting a sensor or 
signal source to your meter. 
8.4.2 Scaling Your Meter 
Scaling your meter without a sensor or a signal source connected is easily accomplished by using one of the 
calculated scale factors as shown in Figure 8-1. 
CF 67  27  M1291/N/0403 11279ML-02 Rev. A 