enter “–00020” in Reading Offset (“RDG OF”) and if your zero input shows a negative “–20” you would enter a 
positive 000020 in Reading Offset (“RDG OF”). 
PRESS THEN PRESS 
(TO CHANGE 
IF REQUIRED) 
UNTIL 
DISPLAY 
SHOWS 
COMMENTS 
‘MIN’  ‘MAX’  “XXXXXX”  Use the ‘MIN’ button to move to each digit and the ‘MAX’ button 
to change the flashing digits value and enter your zero offset 
number. 
‘MENU’ 
“STORED”  Press the ‘MENU’ then button to store your “IN CNF” selection. 
The display will momentarily show “STORED” then “IN CNF”. 
‘RESET’  ‘RESET’  “RESET2” then 
the Measured 
Value 
Press the ‘RESET’ button two times. The display will momentarily 
show “RESET2” and then will display the currently measured 
values. 
 8.4.4 Entering Zero Offset Numbers 
The RDG OF (Reading Offset) menu item should be used if the meter shows a nonzero reading with zero input. 
The offset value zeroes the display by canceling out the nonzero reading. 
If your meter displays a positive reading at zero input, you must enter a negative offset value. If your meter displays 
a negative reading at zero input, you must enter a positive offset value. 
If you are using an active decimal point (RDG.2=0), your offset value will be the negative of the display reading at 
zero point. 
If you are using the more common independent decimal point (the factory setting, RDG.2=1), follow these 
instructions to convert the display reading to the appropriate offset value: 
1.  Note the display reading at zero input, ignoring the decimal point. This reading represents the count value - the 
number of whole counts that need to be offset. 
2.  Shift the count value to the left side of the decimal point. 
3.  If the count value is positive, make it negative by replacing the leading digit (the left-most digit) with a minus 
sign. If the count value is negative, make it positive by replacing the negative sign with a zero. 
Example 1: Your meter displays 000.003 when the input is zero. The count value is 000003. Shift this value to the 
left side of the decimal point: 003.000. Change the leading zero to a minus sign: 
–03.000. This is the “converted offset value” you will use for configuring RDF OF. 
Example 2: Your meter displays –00.003 when the input is zero. The count value is –00003. Shift this value to the 
left side of the decimal point: –03.000. Change the leading minus sign to a zero: 003.000. This is the “converted 
offset value” you will use for configuring RDG OF. 
If the nonzero reading is fluctuating between two numbers, convert the smaller count value to the offset value, then 
add a 5 just right of the decimal point. This adds half a count to the offset. 
For example, the display is fluctuating between 00.0001 and 00.0002. Calculate the offset using the 000001 count 
value. The converted offset value is -1.0000. Add a 5 to the right of the decimal point: the final offset value is -
1.5000. 
PRESS  THEN PRESS 
(TO CHANGE 
IF REQUIRED) 
UNTIL 
DISPLAY 
SHOWS 
COMMENTS 
‘MENU’    “RDG OF”  Press the ‘MENU’ button until the display shows “RDG OF”. 
‘MIN’    “000000.”  Press the ‘MIN’ button and the display will show the last offset 
entered. 
‘MIN’  ‘MAX’  “XXXXXX”  Use the ‘MIN’ to move to each digit and the ‘MAX’ button to 
change the flashing digits value and enter your zero offset 
number. 
CF 67  30  M1291/N/0403 11279ML-02 Rev. A 