Instruction Manual
•  multipoint or point-to-point mode [BUS.4] 
•  (if in point-to-point mode) select whether to communicate continuously or on command [BUS.5] 
•  whether a message character is used in handshake or continuous mode [BUS.6] 
•  whether or not you have installed the RS-485 board [BUS.7] 
•  whether or not to enable the external print command at P2-11. 
MENU BUTTON 
MAIN MENU 
MIN/MAX/MENU * 
BUTTON SUB MENU 
DESCRIPTION 
BUS FORMAT 
Check sum with reading: 
“BUS.1=0”  Excluded 
“BUS.1=1”  Included 
Line feed following all <CR>’s: 
“BUS.2=0”  Excluded 
“BUS.2=1”  Included 
Response from the meter (echo): 
“BUS.3=0”  No 
“BUS.3=1”  Yes 
Point to Point mode or Multipoint mode: 
“BUS.4=0”  Pt-Pt 
“BUS.4=1”  Multi-Pt 
Point-to-Point mode only: 
“BUS.5=0”  Continuous 
“BUS.5=1”
  On Command 
  Handshake if continuous mode: 
“BUS.6=0” Message 
“BUS.6=1” Character 
  RS-485 Board installed: 
“BUS.7=0” Not installed 
“BUS.7=1” Installed 
  Print Command at P2-11: 
“BUS.8=0” Disabled 
“BUS FT”: 
“BUS.8=1”
 Enabled 
* The ‘MIN’ button allows you to sequence through BUS.1, BUS.2, BUS.3, BUS.4, BUS.5, BUS.6, BUS.7 and 
BUS.8. 
The ‘MAX’ button allows you to select the “0” or “1” state for each “BUS” condition. 
The ‘MENU’ button stores the selected values for each “BUS” condition changed and advances the meter to the 
next configuration (“SERCNT”). 
Every underlined “0” or “1” state is the factory preset value. 
9.2.23 SERCNT (Serial Count) 
Serial count is used to program the number of readings the meter must take (programmable from “00001” to 
“59,999”) between transmissions of data via the optional RS-232 or RS-485 serial communications board. 
Once you are done with the changes, press the ‘MENU’ button to display the normal operating display. The meter 
advances to run mode showing the currently measured values. Pressing the ‘RESET’ button two times allows you 
to return to the run mode. 