Instruction Manual
TO CORRECT: Reduce the “CNT BY” count by value to 001. If you have an active decimal point selected, move 
the decimal point one or more positions to the right. 
UOM.OVF 
POSSIBLE CAUSE: Selection of unit of measure displayed (“RDG.6=0”) moves the display reading 
one place to the left causing a numerical overflow. 
TO CORRECT: If you have an active decimal point, move he decimal point position one or more positions to the 
right. 
SERIAL 
POSSIBLE CAUSE: A configuring change has been attempted via the front panel buttons while the serial 
communications port is actively communicating with the meter. 
TO CORRECT: Either lockout the front panel buttons by removing the jumpers S3A & S3C or disconnect the serial 
communications option. 
18.0 SPECIFICATIONS 
18.1 Current Input 
INPUT RANGES 
(+10%): 
0-20 mA and 4-20 mA 
RESOLUTION: 1µA 
MAXIMUM INPUT:  200 mA 
INPUT OHMS:  5 
SENSOR EXCITATION:  10 V, to 30 mA 
24 V, to 25 mA 
18.2 Voltage Input 
INPUT RANGES:   
 UNIPOLAR:  100 mV 1 V 10 V 100 V 
 BIPOLAR:  ±50 mV ±0.5 V ±5 V ±50 V 
RESOLUTION: 1 µV 10 µV 100 µV 1 mV 
MAX INPUT:  70 Vp 350 Vp 350 Vp 300 Vp 
BIAS AMPS: 
SENSOR EXCITATION:  10 V, to 30 mA 
24 V, to 25 mA 
18.3 Potentiometer Input 
RESOLUTION: 
1.5 to 11 Vdc, to 60 mA Max 24 V, to 25 mA Max 
18.4 General 
SCALE:  +0.000001 to +500000 or –0.0001 to –99999. 
Zero to +999999 
NMR: 
INPUT OHMS:  1G 1M 1M  1M 
50 pA 5 pA 1 pA 1 pA 
INPUT RANGES:  1 V or 10 V 
programmed to 0.001% 
SENSOR EXCITATION: 
OFFSET: 
POLARITY: Automatic 
NOISE REJECTION 
60 dB, 50 or 60 Hz, + selected filter 
CMR:  120 dB 
CMV:  Dielectric strength to 1500 V transient per 260 VRMS or DC working voltage. 
ACCURACY at 25 C: 
MAX ERROR:  ±0.005% of reading 
CF 67  70  M1291/N/0403 11279ML-02 Rev. A 