Instruction Manual
1  Rear Protective Cover with Screw 
1 20-Socket Ribbon Connector 
(P2 Connector) 
2 Panel-Mounting Gaskets 
(1 Spare) 
1  Strain Gauge Owner’s Guide   
1  Quick Start Manual   
Other items may also be in the box depending on the options ordered. Refer to specific options described 
previously. 
3.0 SAFETY CONSIDERATIONS 
This device is marked with the international caution symbol. It is important to read this manual before installing or 
commissioning this device as it contains important information relating to Safety and EMC (Electromagnetic 
Compatibility). 
Unpacking and Inspection 
Unpack the instrument and inspect for obvious shipping damage. Do not attempt to operate if the unit is damaged. 
This instrument is a panel mount device protected in accordance with Class I of EN 61010 (115/230 AC power 
connections), Class II for the DC power option (10-32 Vdc). Installation of this instrument should be done by 
qualified personnel. In order to ensure safe operation, the following instructions should be followed. 
This instrument has no power-on switch. An external switch or circuit breaker shall be included in the building 
installation as a disconnecting device. It shall be marked to indicate this function and it shall be in close proximity to 
the equipment with easy reach of the operator. The switch or circuit breaker shall not interrupt the Protective 
Conductor (Earth wire) and it shall meet the relevant requirements of IEC 947-1 and IEC 947-3. The switch shall 
not be incorporated in the mains supply cord. 
Furthermore, to provide protection against excessive energy being drawn from the mains supply in case of a fault in 
the equipment, and overcurrent protection device shall be installed. 
•  The protective conductor must be connected for safety reasons. Check that the power cable has the proper 
Earth wire, and it is properly connected. It is not safe to operate this unit without the protective conductor 
terminal connected. 
•  Do not exceed voltage rating on the label located on the top of the instrument housing. 
•  Always disconnect power before changing signal and power connections. 
•  Do not use this instrument on a workbench without its case for safety reasons. 
•  Do not operate this instrument in flammable or explosive atmospheres. 
•  Do not expose this instrument to rain or moisture. 
CF 67  3  M1291/N/0403 11279ML-02 Rev. A 