Items you can change: 
Power On Auto Zero   Here you can determine whether the device performs 
   automatic zero when it is powered on. 
   Enter zero to disable this function. 
   If you enter a non-zero value then when the handheld 
   is first turned on it checks the value read from the 
   acquisition module. If this falls within ± of this value 
   then the display will be altered so this reads zero. 
   In its simplest form you could select two for a linear 
   calibration. For more complex calibrations which 
   include linearization select three to nine points. 
Zero Indication Band   Using this setting you can mask tiny changes in input 
   after you press the Tare button. 
   Entering zero will disable this function. 
   Entering a non-zero value will provide a band within 
   which the display will always read zero. 
   Once the reading exceeds this value the real weight  
   will be displayed as no taring is taking place. 