User guide
Model 3i Digital Force/Torque Indicator    User’s Guide 
15
 If failed:   
5.  The following screen appears after the offsets have been calculated: 
Attach weight fixtures (brackets, hooks, etc), as required. Do not yet attach any weights or apply 
any calibration loads. Then press ENTER. 
6.  The display will appear as follows: 
Optionally exercise the sensor several times (at full scale, if possible), then press ENTER. 
7.  The display will appear as follows: 
Apply a weight equal to the full scale of the instrument, then press ENTER. 
8.  After displaying “PLEASE WAIT…” the display will appear as follows: 
CALIBRATION 
COMPRESSION 
GAIN ADJUST 
APPLY FULL SCALE LOAD 
10.000 LBF +/-20% 
THEN PRESS ENTER 
CALIBRATION 
COMPRESSION 
Attach necessary 
weight fixtures. 
THEN PRESS ENTER 
CALIBRATION 
COMPRESSION 
Optionally exercise 
load cell a few times. 
THEN PRESS ENTER 
CALIBRATION 
OFFSET 
 Sen.Offset Adj.Failed 
 Ana.Offset Adj.Failed 
CALIBRATION 
OFFSET 
Sen.Offset Adj.Passed 
Ana.Offset Adj.Passed 
CALIBRATION 
OFFSET 
Please wait… 