Owner's manual
Model 7i Digital Force / Torque Indicator    User’s Guide 
11
 – the displayed value is greater than the upper 
set point (NO GO HIGH) 
  – the displayed value is between the set points 
(GO) 
 – the displayed value is less than the lower set 
point (NO GO LOW) 
Note: Set point indicators and outputs reference the displayed reading, not necessarily the current live 
load. 
6.3 Set Point Outputs Schematic Diagram 
6.4 Using Set Points to Control a Mark-10 Motorized Test Stand 
When using set points to stop/cycle Mark-10 motorized test stands, the upper and lower set points must 
be set to opposite measuring directions. Both set points must be set, even if the intended application is 
to stop/cycle at only one of the set points. The opposite set point should be a value sufficiently large that it 
does not get triggered during the course of the test. 
For certain Mark-10 test stands, the upper and lower set point directions are reversed. 
7 BREAK DETECTION 
The break detection function identifies when a sample has broken, clicked, slipped, or otherwise reached 
a peak load and then fell by a specified percentage drop. Upon detection of the break, the indicator can 
perform several automatic functions, depending on the mode in which break detection is enabled, as 
follows: 
1.  Transmit the peak reading (Auto Output). 
2.  Save the peak value to memory (Auto Storage). 
3.  Zero the primary and peak readings (Auto Zero). 
4.  Toggle a pin (for example, to stop crosshead movement on a Mark-10 motorized test stand). 
Break detection functions and settings are configured from a central location, and apply to any mode in 
which it is enabled. Refer to the Operating Modes section for details on configuring each mode. 
7.1 Configuration 










