Manual

The display now shows the different submenus you may choose:
5.1 Calibration Data
Calib-Data
With the buttons under the arrows you may choose the function you want and with
-button the submenu is opened. You will find the different submenus as follows:
HYGROCONTROL Manual for Type 86 Ex Page
5. Operating the Instrument with Display-Option
From normal display of Humidity and Temperature you get into the menus for
Configuration and Calibration by pressing one of the 4 buttons on the display-
board.
26
-- SETTINGS --
List Calibration Data
List Configurat. Data
Calibrate (PW)
Cofigure (PW)
Display Setup
4
4
F freq: 51732
R freq: 51217
F freq: 51732
R freq: 51217
= brings the next value into the display
= also brings the next value into the display
= Stops display of calibration data
= encreasing as long as sensors frequency is detected
= actual Frequency of the Humidity sensor
= actual Frequency of the Reference
Setup: 0.0%RH
F diff: 1395
Rec.#: 0 ReCal:101
Any pressing of the “arrow”-buttons brings the next
calibration point and the corresponding difference
frequency into the display
Rec.# = Number of values in the memory,
20 and 21 are first 2 values of factory calibration
ReCal: 101 means: 1. Digit = deleted by customer
2. and 3. Digit = 1x recalibrated by customer
This menu shows stored values of calibration data, but values cannot be changed!
= letters are white on back background
and this is the submenu choosen.
(PW) = for these submenus a Password is neccessary.
Calib-Data