Manual

Relays
31
Time Controlled
The relay can be automatically turned on and off based on a specific time and day. For example, it can
be set to turn on automatically every weekday at 7:00AM and to turn off at 5:00PM. This is an added
benefit for the customers to utilize the instrument as a small local control system.
Set relay X to time and day controlled
Go to Relays > Channel X > Variable
Select Time
Select the day relay turning on/off
Go to Relays > Time > Mode
Select the one of the following:
Everyday
Every weekday
Weekly
Monthly
Select the day when relay operates on weekly or
monthly basis
Go to Relays > Time > Day
Change the time when relay turns on/off
Go to Relays > Time > Time On
Set the time relay turning on in 24-hour format
Go to Relays > Time > Time Off
Set the time relay turning off in 24-hour format
Remote Controlled
Individual relay can be controlled remotely via the Modbus communication. Once the relay is set to
operate on communication command, it can be turned on/off by changing the Modbus coil registers.
This allows the DCS system to control another device via the XTDL-HT™ without setting up a Remote
Terminal Unit.
Set relay X to remote controlled
Go to Relays > Channel X > Variable
Select Communication