User's Manual

Table Of Contents
Sampling line maintenance (Permapure)
Do not bend, squash or deform the sampling line. Any “kink” in the sample line will reduce the internal lumen of the line and affect
accuracy of measurement.
Do not keep the sampling line open to the atmosphere, particularly in crowded or smoky environments. Keep the sampling line in sealed
plastic bag in a dark cool and dry place.
If saliva enters the tube it should be replaced immediately.
Periodically grease the O-ring on the connector to ease fitting to optical flowmeter.
Replace the sampling line every 100 exercise tests or 200 PFT test or every 6 months. In any case, sampling line will become discoloured
(brown) with age and may cause calibration to fail.
Replace the sampling line if the Autocheck failed.
Note: ALWAYS replace sample line as the first step in troubleshooting a failed gas calibration.
60 - System maintenance - Sampling line maintenance (Permapure)