Chapter 2 System Setup
2.2.3 Connecting the ACIN
The POC-621-01 SERIES can only be powered by AC internal power (FSP Model
no. FSP150M-K24-18).
Be sure to always handle the power cords by holding the plug ends only.
Follow these procedures in order:
1. Connect the female end of the power cord to AC in.
2. Connect the 3-pin male plug of the power cord to an electrical outlet.
2.2.4
Connecting the Ground pin
1. System ready and find the Equipotential Terminal on rear side of POC.
An Equipotential Terminal is provide to optionally connect to a hospital ground/
earth system
Figure 2.6 POC-621 Equipotential Terminal Pin