For transmitters rated at a maximum output power not listed above, the recommended sep-
aration distance d in metres (m) can be estimated using the equation applicable to the fre-
quency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Guidance and manufacturer’s declaration – electromagnetic immunity
POC-621 SERIES is intended for use in the electromagnetic environment specified below.
The customer or the user of the model POC-621 SERIES should assure that it is used in
such an environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environ-
ment –guidance
Electrostatic dis-
charge (ESD)
IEC 61000-4-2
8 kV contact
15 kV air
8 kV contact
15 kV air
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for input/out-
put lines
2 kV for pow-
ersupply lines
1 kV for input/
output lines
Main power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
1 kV line(s) to
line(s)
2 kV line(s) to
earth
1 kV line(s) to
line(s)
2 kV line(s) to
earth
Main power quality should be
that of a typical commercial or
hospital environment.
Interruptions and
voltage varia-
tions on power
supply input lines
IEC 61000-4-11
Voltage Dips
0% reduction for
0.5 / 1 cycle at
50Hz
70% reduction for
25 / 30 cycle at 50/
60Hz
Voltage Interrup-
tions
0% reduction for
250 / 300 cycle at
50/60Hz
Voltage Dips
0% reduction
for 0.5 / 1
cycle at 50Hz
70% reduction
for 25 / 30
cycle at 50/
60Hz
Voltage Inter-
ruptions
0% reduction
for 250 / 300
cycle at 50/
60Hz
Main power quality should be
that of a typical commercial or
hospital environment. If the
user of the model POC-621
Series requires continued
operation during power
mains interruptions, it is rec-
ommended that the model
POC-621 Series be powered
from an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m 30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commer-
cial or hospital environment.
NOTE UT is the A.C. mains voltage prior to application of the test level.