Tus neeg siv phau ntawv

Table Of Contents
POC-621 Series User Manual 24
Guidance and manufacturer’s declaration – electromagnetic immunity
The model POC-621 SERIES is intended for use in the electromagnetic environment spec-
ified below. The customer or the user of the model POC-621 SERIES should assure that it
is used in such an environment.
Immunity
test
IEC 60601 test
level
Compliance
level
Electromagnetic environment –
guidance
Conducted
RF
IEC 61000-4-
6
Radiated RF
IEC 61000-4-
3
3 Vrms
150 kHz to 80
MHz
3 V/m
80 MHz to 2,5
GHz
Vrms
V/m
Portable and mobile RF communi-
cations equipment should be used
no closer to any part of the model
POC-621 Series, including cables,
than the recommended separation
distance calculated from the equa-
tion applicable to the frequency of
the transmitter.
Recommended separation dis-
tance
d = 1,2
d = 1,2 80 MHz to 800 MHz
d = 2,3 800 MHz to 2,5 GHz
where P is the maximum output
power rating of the transmitter in
watts (W) according to the transmit-
ter manufacturer and d is the rec-
ommended separation distance in
metres (m).
Field strengths from fixed RF trans-
mitters, as determined by an elec-
tromagnetic site survey,
a
should be
less than the compliance level in
each frequency range.
b
Interference may occur in the vicin-
ity of equipment marked with the fol-
lowing symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption
and reflection from structures, objects and people.
P
P
P