Section 4: Service Mode
4.3.1 Default Alarm Limits
The service technician can change the default alarm limits for either
patient mode (Adult or Neonatal) as described in Table 4-3: "Changing
Default Alarm Limits".
The monitor has the following alarms with adjustable level settings:
•EtCO
2 high and low levels (mmHg)
•FiCO
2 high level (mmHg)
•SpO
2 high and low levels (%)
Pulse rate high and low levels (bpm)
Warning: Changing default settings will permanently change the
user’s start-up value and should be verified with the responsible
clinical personnel.
Table 4-3: Changing Default Alarm Limits
Objective Action Response
To access the Default
Alarm Limits menu
after entering last
password digit
short press
To change the patient
mode
short press
To access a new
parameter
short press
To change the value short press