Section 4: Service Mode
4.3.2 Default Settings
The service technician can change two default settings (refer to Table 4-4:
"Changing Default Settings"):
•3 Min Alert
When the 3 Min Alert is ON, the unit will beep every three minutes
to remind the user that an alarm (or all alarms) are set to OFF.
BTPS (body temperature, pressure, saturation)
When BTPS is ON, the CO
2 value is corrected for this factor
Pulse Tone
Pulse Tone default is ON. When changed to OFF, the Pulse Tone
default is silent (as indicated on screen).
.
Table 4-4: Changing Default Settings
Objective Action Response
To access the Default
Settings menu (after
accessing the Default
Alarm Limits menu)
long press
To change the setting short press
To access BTPS
parameter (and save
the changed setting for
3 Min Alert)
short press
To reset and return to
measuring mode
(automatically erasing
event messages
memory)
long press
To exit service mode
and save event
message
Turn off the
monitor.