Product Manual

Section 7: Disassembly Guide
7-10
7.6.2.2 Reassembly of the lines to Solenoid
Refer to Figure 7-10: Lines to Solenoid (Zero line, Input line 1,
Input line 2) and Figure 7-11: Zero line
1. Connect the new Input line1 to the line 1 solenoid connector.
2. Connect the new Input line 2 to the line 2 solenoid connector.
3. Bring the Input line 2 tubing over Input line 1, and connect to
the sensor connector.
4. Lead the Input line 1 under the FRS cables (connected to the
FilterLine input connector, not viewed in Figure 7-10) and
connect to the FilterLine input connector.
5. Connect the Zero line to the Zero line solenoid connector. Place
the scrubber close to the pump under the coiled Main line
tubing (not viewed in Figure 7-11: Zero line) around the
pressure sensor.
Figure 7-11: Zero
line